TELEDYNE Model 9110AH NO
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
11
OPTIC_TEST
12
BROWNOUT_RESET
13
CONV_HEATER
14
O3GEN_STATUS
15
ST_LOW_SPAN_CAL
16
AUTOZERO_VALVE
17
ST_RCELL_PRESS
18
PREAMP_RANGE_HI
19
ST_CONC_ALARM
20
ST_ZERO_CAL
21
ST_SPAN_CAL
(table continued)
Analyzer Operator Manual, 01620, Rev. F
X
Control
Description
YES
Turns on optic test bit in preamp. Should be used for
troubleshooting Preamp logic lines. We recommend
you use the OT button in the DIAG menu to operate
optic test.
YES
Brownout reset works in conjunction with
DSP_BROWNOUT. When DSP_BROWNOUT is set
the CPU sends a signal to reset the display and clear the
DSP_BROWNOUT.
NO
Shows the status of the converter heater. This has the
same function as the LED in the PSM.
YES
Switches power to the ozone generator. Can be used to
force the ozone generator to turn on.
YES
Status Bit - state of the low concentration span valve
Logic high = valve open to admin span gas
Logic low = valve closed
YES
Switches the AutoZero valve. Use this bit to test the
valve function.
YES
Status Bit - Reaction Cell Pressure alarm
Logic high = pressure out of acceptable range
Logic low = pressure inside acceptable range
YES
Switches the preamp hardware range. Standard ranges
are 500 and 5,000 ppm. Logic high = 5,000 ppm; logic
low = 500 ppm. M9110AH will reset range to correct
value based on user set range value.
YES
YES
Status Bit - Zero Calibration mode
Logic high = M9110AH in Zero cal mode
Logic low = Not in Zero cal mode
YES
Status Bit - Span Calibration mode
Logic high = M9110AH in Span cal mode
Logic low = Not in Span cal mode
9-13