TELEDYNE Model 9110AH NO
Enter the expected NO
Table 7-3: Enter Expected Span Gas Concentrations Procedure
Step Number
1.
2.
3.
4.
5.
6.
If desired, compensation for converter efficiency (CE) can be included in the NO
concentration calculation. The CE must be entered prior to calibration. Refer to Section 7.9 for
the CE procedure
Analyzer Operator Manual, 01620, Rev. F
X
and NO span gas concentrations:
x
Action
Comment
Check "Before
Check "Before Calibration" Section 7 before proceeding.
Calibration"
Press
This key sequence causes the M9110AH to prompt for the
CAL-CONC-NOX
expected NO
Enter the NO
under each digit until the expected value is set. This menu can
also be entered from CALS or CALZ.
Press ENTR
Pressing ENTR stores the NO span value and returns the
prompt to the CONC menu.
CAL-CONC-NO
This key sequence causes the M9110AH to prompt for the
expected NO concentration.
Enter the NO span concentration value by pressing the key
under each digit until the expected value is set. This menu can
also be entered from CALS or CALZ
Press ENTR
Pressing ENTR stores the NO span value and returns the
prompt to the CONC menu.
Press EXIT
Returns instrument to SAMPLE mode.
7-4