Teledyne 9110AH Instruction Manual page 131

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
33
PMT_TEMP
34
DCPS_VOLTAGE
35
RCELL_PRESS
36
OZONE_FLOW
37
CONV_TEMP
38
HVPS_VOLTAGE
39
DAC_CHAN_0
40
DAC_CHAN_1
41
DAC_CHAN_2
42
DAC_CHAN_3
43
NOX_CONC
44
NO_CONC
45
NO2_CONC
46
TEST_OUTPUT
Analyzer Operator Manual, 01620, Rev. F
X
Control
Description
NO
PMT cold block temperature. Typically 4300 mV for 7
C.
NO
DC power supply composite voltage output. Typically
2500 mV
NO
Reaction Cell Pressure in mV. Typically 1270 mV for
5" Hg-A at sea level. Is an absolute pressure so higher
values means higher absolute pressures.
NO
Ozone flowmeter voltage. Typically 3200 mV at
250 cc/min.
NO
Converter temp. Typically XXXX mV at 700 C
NO
HVPS programming voltage. Output of HVPS is
1000x value present. 700 mV = 700 VDC output.
NO
Output of NO
NO
Output of NO channel in mV.
NO
Output of NO
NO
Test Channel output. Also second NO
YES
NO
DAC programming voltage. The following 3
x
signals can be set to output specific voltages to each
DAC. Use in conjunction to ANALOG OUTPUT test
to check each DAC output channel. The value keyed in
should appear on the appropriate analog output
channel. This value overrides data being written from
the analyzer. Value reverts to instrument output when
function is exited.
YES
NO DAC programming voltage. See above for
description.
YES
NO
DAC programming voltage. See above for
2
description.
NO
Displays voltage of selected analog signal
9-15
channel in mV.
x
channel in mV.
2
output.
x