Maintenance; Maintenance Schedule; Table 8-1: Preventativem - Teledyne 9110AH Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO

8 MAINTENANCE

8.1 Maintenance Schedule

The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Table 8-1: Preventative Maintenance Schedule
Item
TEST functions
Zero/Span Calibration
Zero/Span checks
Particulate Filter
Filter for PermaPure Drier
Filter for Ozone Scrubber
Drierite Canister
Reaction cell window
Ozone Flow
Sample Flow
Converter
Pneumatic Lines
Factory Calibration
Leak Check
Rx Cell O-rings
O-rings
Analyzer Operator Manual, 01620, Rev. F
X
NOTE
Maintenance Interval
6 - 12 month intervals
Annually or after repairs
Daily
Weekly as needed
Replace every 24 months
Replace every 24 months
Replace when Drierite changes
from blue to red.
Clean annually or as necessary
Check TEST function every year
Check TEST function every year
Check efficiency every 6 months
Examine every 12 months, clean
if necessary
Calibrate each year or after
repairs
Check every year or after other
maintenance.
Replace every 12 months
Replace every 12 months
8-1
Reference Section
Table 9-1
Section 7
Section 7, Table 6-2
Figure 8-1
Figure 9-7
Figure 4-2
Section 9.3.6
Section 9.3.8, Figure 8-3
Figure 9-7, Section 9.3.7
Figure 8-4, Figure 8-5
Section 9.3.7
Figure 8-2, Section 7.9
Figure 8-5, Figure 8-6
Section 9.1.6
Section 8.6, 9.3 10
Figure 8-3
Figure 9-9, Figure 9-10