Teledyne 9110AH Instruction Manual page 130

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
22
ST_FLOW_ALARM
23
ST_TEMP_ALARM
24
ST_DIAG_MODE
25
ST_POWER_OK
26
ST_SYSTEM_OK
27
ST_HIGH_RANGE
28
PMT_SIGNAL
29
SAMPLE_PRES
30
RCELL_TEMP
31
BOX_TEMP
32
BLOCK_TEMP
(table continued)
Analyzer Operator Manual, 01620, Rev. F
X
Control
Description
YES
Status Bit - Flow alarm
Logic High = Sample/Ozone flow out of spec
Logic Low = Flows within spec
YES
Status Bit - Temperature alarm
Logic High = Rxcell, Conv, Box temps out of spec
Logic Low = Temps within spec
YES
Status Bit - In Diagnostic mode
Logic High = M9110AH in Diagnostic mode
Logic Low = Not in Diag mode
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
YES
Status Bit - System OK
Logic High = No instrument warnings present
Logic Low = 1 or more alarms present
YES
Status Bit - Autorange High Range
Logic High = M9110AH in high range of autorange
mode
Logic Low = M9110AH in low range of autorange
mode
NO
Current PMT voltage. Same as PMT voltage in TEST
menu. Bi-polar, typically in 0-5000 mV range. A
constant value of 5000 mV indicates offscale.
NO
Sample pressure in mV. Typical sea level value = 4300
mV for 29.9" Hg-A.
NO
Reaction Cell temperature. Typically 3500 mV for
50 C.
NO
Box Temperature. Typically 1800 mV for 25 C
NO
Temperature of the sample flow control orifice.
Typically 3500 mV for 50 C.
9-14