TELEDYNE Model 9110AH NO
Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Analyzer Operator Manual, 01620, Rev. F
X
Action
Comment
Input Zero gas
Allow Zero gas to enter the sample port on the rear of the
instrument.
Press CAL
The M9110AH enters the calibrate mode from sample mode.
Wait 10 min
Wait for reading to stabilize at the zero value. If you wait less
than 10 minutes the final zero value may drift.
Press ZERO
The ZERO button will be displayed.
Press ENTR
Pressing ENTR actually changes the equations and zeros the
instrument.
Press EXIT, input
M9110AH returns to the CAL menu. Now switch gas streams
Span gas
to span gas.
Wait 10 min
Wait for reading to stabilize at the span value.
Press SPAN
The SPAN button should be displayed. If there is no SPAN
button check the Troubleshooting Section 9.2.8 for
instructions on how to proceed. In certain circumstances at
low span gas concentrations both the ZERO and SPAN
buttons will appear.
Press ENTR
Pressing ENTR actually changes the equations so that the
concentration displayed is the same as the expected span
concentration you entered above, thus spanning the instrument.
Press EXIT
Pressing EXIT returns the instrument to SAMPLE mode.
2-8