TELEDYNE Model 9110AH NO

5.6 Status Output

The status output is an option that reports Analyzer conditions via contact closures on the rear
panel. The closures are available on a 50 pin connector on the rear panel. The contacts are NPN
transistors which can pass 50 ma of direct current. The pin assignments are listed in Table 5-8
below.
Table 5-8: Status Output Pin Assignments
Output #
1
2
3
4
5
6
7
8
9
10
11
12
The Status Board schematic can be found in the Appendix Drawing 01087.
5.7 RS-232 Interface
The RS-232 communications protocol allows the instrument to be connected to a wide variety
of computer based equipment. The interface provides two basic functions in the M9110AH.
1. First is a comprehensive command interface for operating and diagnosing the analyzer. The
command set can be used interactively or operated via stored program. This interface has,
in fact, more capabilities than the front panel keyboard.
2. The interface can provide an audit trail of analyzer events. In this function the port sends
out messages about instrument events like calibration or warning messages. If these
messages are captured on a printer or remote computer, they provide a continuous audit
trail of the analyzers operation and status.
Analyzer Operator Manual, 01620, Rev. F
X
Pin #
Definition
1,2
ZERO CAL
3,4
SPAN CAL
5,6
FLOW ALARM
7,8
TEMP ALARM
9,10
DIAG MODE
11,12
POWER OK
13,14
CONC ALARM
27,28
LOW SPAN CAL
SPARE
19,20
AUTORANGE - HI
21,22
SYSTEM OK
23,24
RX CELL PRESS
Condition
CLOSED IN ZERO CAL
CLOSED IN HIGH SPAN CALIBRATION
CLOSED IF FLOW WARNING
CLOSED IF ANY TEMP WARNING
CLOSED IN DIAG MODE
CLOSED IF SYSTEM POWER OK
CLOSED IF CONCENTRATION WARNING
CLOSED IN LOW SPAN CALIBRATION
CLOSED IF IN HIGH RANGE
CLOSED IF NO FAULTS PRESENT
CLOSED IF ABS PRES > 15" HG
5-25