Specifications, Warranty; Specifications - Teledyne 9110AH Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO

3 SPECIFICATIONS, WARRANTY

3.1 Specifications

Operating Modes
Ranges
Noise at zero
Noise at span
Detection Limit(Note 1)
Zero Drift (Note 2)
Zero Drift (Note 2)
Span Drift (Note 2)
Lag Time
Switching Mode
NO
mode
x
Response Time
Switching Mode
NO
mode
x
Sample Flow Rate
Linearity
Precision
Temperature Range
Humidity
Temp Coefficient
Voltage Coefficient
Dimensions HxWxD
Weight, Analyzer
Weight, Pump Pack
Power, Analyzer
4
Power, Analyzer
Power, Ext Pump
4
Power, Ext Pump
Environmental
Analog Resolution
Recorder Output
Current Loop Option
Status
Measurement Units
1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Lag & response times longer for external converter option.
4. Electrical ratings for CE Mark compliance.
Analyzer Operator Manual, 01620, Rev. F
X
NO
NO/
switching mode, NO only mode,
x
In 1 ppm increments from 5 ppm to 5,000 ppm
Single range, independent ranges or autoranging
0.02 ppm RMS
<0.2% of reading RMS above 20 ppm
0.04 ppm RMS
<0.2% full scale/24 hours
<0.4% full scale/7 days
<1% FS/24 hours
20 sec (Note 3)
4 sec (Note 3)
95% in < 40 sec (Note 3)
95% in < 10 sec (Note 3)
290 10 cc/min (Including bypass)
1% of full scale
0.5% of reading
o
5-40
C
0-95% RH non-condensing
o
< 0.1% per
C
< 0.1% per V
7"x17"x23.6" (18 cm x 43 cm x 61 cm)
43 lbs (20 kg)
16 lbs (7 kg)
100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts
230 V~ 50 Hz, 2.5A
110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts
230 V~ 50 Hz, 2.5A
Installation Category (Over-voltage Category) II
Pollution Degree 2
1 part in 2048 of selected voltage or current range
0-100 mV, 0-1, 5, 10v, bipolar
4-20ma isolated
12 Status Outputs from opto-isolator
3
ppm, mg/m
3-1
NO
only mode
x

Advertisement