Table 9-1: Test Functions (Continued)
Test Function
NORM PMT
AZERO
HVPS
DCPS
RCELL TEMP
BOX TEMP
PMT TEMP
BLOCK TEMP
(table continued)
Analyzer Operator Manual, 01620, Rev. F
X
Factory Set-Up
Comment
0-5000 mV
Normalized PMT voltage. Takes raw PMT voltage and applies
temperature and pressure corrections. Use NORM PMT when doing
the factory calibration procedure. A slope of 1.0 can be achieved for
high altitude sites and pumps that are significantly down from their
maximum vacuum.
-20-100 mV Check
AZERO is the current value of the AutoZero circuit reading. The
value in
number typically should around zero. The value may be temporarily
Table 2-1
outside this range if the PMT has been recently been exposed to light
or the instrument has been turned off for several days. If the value
stays outside this range, check the AutoZero and NO/NO
cross port leak. Use Diagnostic mode to manually check the
AutoZero valve for correct operation.
400-600 VDC
This represents the scaled-up HVPS programming voltage to the
HVPS. The design of the HVPS precludes taking a single reading that
indicates the health of the supply. Refer to the HVPS
Troubleshooting Section 9.3.8.5 for a procedure for testing the
HVPS. This TEST function is used primarily to set the HVPS voltage
value. The supply will function over a voltage range of 375 to 1000
VDC. A value in the 400 to 600 volt range is common for the
M9110AH.
2500
200 mV
DCPS is a composite of the +5 and
arbitrarily set at 2500
voltages in the supply is not working. Check the procedures for
diagnosing the Power Supply Module.
50
2
The reaction cell temperature is controlled to 50 C
computer. It should only read other values when the instrument is
warming up. If the value is outside the acceptable range, go to the
procedure for diagnosing the Reaction cell temp supply. The alarm
limits are less than 45 C and greater than 55 C.
8-48 C
The Box Temp is read from a thermistor on the Status/Temp board
(01086). It should usually read about 5 C above room temp. The
M9110AH is designed to operate from 5 to 40 C ambient. Therefore
the box temperature should be in the range of about 10 to 50 C.
Temperatures outside this range will cause premature failures of
components, and poor data quality. Warning limits are < 8 C and >
48 C.
7
1 C
The PMT detector is very temperature sensitive. The PMT
temperature should always be 7 C, except at power-up. Temperatures
more than
circuit. See Section 9.3.8.4 for PMT cooler diagnostic and
troubleshooting. Warning limits are < 5 C and >
14 C.
50 C
2
Temperature of the orifice block that houses the sample flow orifice.
Should be 50 C except when the instrument is warming up.
9-6
15 VDC supplies. It has been
200 mV. If it is not in this range one of the
1 C from the set point indicate problems with the cooler
valve for a
x
2 C by the