5.7.3 TEST Commands and Messages

Table 5-15: RS-232 Test Messages
Name
1
NOXRANGE
1
NORANGE
1
NO2RANGE
2
RANGE
STABIL
FLOW
OFLOW
PMT
NORM PMT
AZERO
HVPS
DCPS
RCTEMP
BOXTEMP
PMTTEMP
BLOCKTEMP
CONVTEMP
RCPRESS
SMPPRESS
NOXSLOPE
NOXOFFSET
NOSLOPE
NOOFFSET
CLKTIME
1
Displayed when independent range is enabled.
2
Displayed when single or AutoRange is enabled.
3
Depends on which units are currently selected.
Analyzer Operator Manual, 01620, Rev. F
X
Message
NOX RNG=xxxxx ppm
3
NO RNG=xxxxx ppm
3
NO2 RNG=xxxxx ppm
3
RANGE=xxxxx ppm
STABIL=xxxx.xx ppm
SAMPLE FL=xxx CC/M
OZONE FL=xxxx CC/M
PMT=xxxxxx MV
NORM_PMT=xxxxxxx MV
AZERO=xxxxxx MV
HVPS=xxxxx V
DCPS=xxxxxx MV
RCELL TEMP=xxx C
BOX TEMP=xxx C
PMT TEMP=xxx C
BLOCK=xxx C
CONV TEMP=xxx C
RCEL=xxx.x IN-HG-A
SAMP=xxx.x IN-HG-A
NOXSLOPE=xxxxx
NOXOFFS=xxxxx
NOSLOPE=xxxxxx
NOOFFS=xxxxxx
TIME=HH:MM:SS
5-33
Description
Indep. range for NO
channel
x
Indep. range for NO channel
Indep. range for NO
channel
2
Analyzer range
Std. Deviation of last 10 min of conc values.
Sample flow rate
Ozone flow rate
PMT output
PMT output adjusted for T/P compensation
AutoZero offset value
High voltage power supply
DC power supply
Reaction cell temperature
Internal box temperature
PMT temperature
Sample flow control block temp
Converter temperature
Rx Cell Pressure
Sample Pressure
NO
Slope Parameter
x
NO
Offset Parameter
x
NO Slope Parameter
NO Offset Parameter
Time of day