TELEDYNE Model 9110AH NO
Step 3 - Review Quality of calibration:
Calibration Quality Check Procedure
Step Number
1.
2.
4.
5.
Step 4 - The M9110AH is now ready to measure sample gas.
Analyzer Operator Manual, 01620, Rev. F
X
Action
Comment
Scroll the TEST
The SLOPE value for NO
function menu until
not in this range, check Section 7.1 or 9. If the SLOPE value is
the NO
SLOPE is
in the acceptable range the instrument will perform optimally.
x
displayed.
Scroll the TEST
The SLOPE value for NO should be 1.0 0.3. If the value is
function menu until
not in this range, check Section 7.1 or 9. If the SLOPE is in the
the NO SLOPE is
acceptable range the instrument will perform optimally.
displayed.
NOTE:
The NO and NO
Scroll the TEST
The M9110AH will display the OFFSET parameter for the
function menu until
NO
the NO
OFFS is
0.0 50 indicates calibration in the optimal range. If the
x
displayed.
OFFSET value is outside this range, check Section 7.1 or 9.1
for procedures to correct the OFFSET value to near zero.
Scroll the TEST
The Instrument will now display the NO OFFSET value. It
function menu until
should also have a value near zero (0.0 50).
the NO OFFS is
displayed.
should be 1.0 0.3. If the value is
x
slopes should be equal within
x
equation. This number should be near zero. A value of
x
2-9
0.3.