TELEDYNE Model 9110AH NO
Table 7-4: Manual Span Calibration Procedure - Span Gas Thru Sample Port
Step Number
1.
2.
3.
4.
5.
6.
7.
Analyzer Operator Manual, 01620, Rev. F
X
Action
Comment
Check "Before
Check "Before Calibration" Section 7 before proceeding.
Calibration"
Press CAL
The M9110AH enters the calibrate mode. NO span gas should
be fed to the sample port.
Select LOW or
If AutoRange is selected, it is necessary to select which range
HIGH
to calibrate. This prompt will not be displayed unless
Autorange or Remote range is selected.
Wait 10 min
Wait for reading to stabilize at span value.
Press SPAN
If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
Press ENTR
Pressing ENTR actually changes the calculation equations and
causes the instrument to read the expected NO and NO
concentrations.
Press EXIT
M9110AH returns to sampling. Immediately after calibration,
data is not added to the DAS averages.
7-5