Series 2600B System SourceMeter® Instrument Reference Manual
Example
smua.source.settling = smua.SETTLE_FAST_POLARITY
Also see
Range
(on page 4-57)
smuX.source.sink
This attribute turns sink mode on or off.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sinkMode = smuX.source.sink
smuX.source.sink = sinkMode
sinkMode
X
Details
This attribute enables or disables sink mode. When sink mode is enabled, it reduces the source limit
inaccuracy that occurs when operating in quadrants II and IV (quadrants I and III will show this source
limit inaccuracy).
Example
smua.source.sink = smua.ENABLE
Also see
Source or sink
2600BS-901-01 Rev. E / January 2019
Affected by
SMU reset
Instrument reset
Recall setup
Sets the sink mode on or off; set to one of the following values:
0 or smuX.DISABLE: Turns off sink mode
1 or smuX.ENABLE: Turns on sink mode
Source-measure unit (SMU) channel (for example, smua.source.sink applies to
SMU channel A)
(on page 7-4)
Section 11: TSP command reference
Selects fast polarity changing for
SMU channel A.
Where saved
Default value
Saved setup
0 (smuX.DISABLE)
Enables sink mode for SMU
channel A.
11-251