Section 11: TSP command reference
display.smuX.measure.func
This attribute specifies the type of measurement that is being displayed.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
func = display.smuX.measure.func
display.smuX.measure.func = func
func
X
Details
Selects the measurement function that is displayed on the front panel: Amps, volts, ohms, or watts.
SMU A and SMU B can be set for different measurement functions.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
Also see
display.smuX.limit.func
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type
TSP-Link accessible
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait() call. This function clears the trigger event detector and discards the
previous history of TRIG key presses.
This attribute also clears the display.trigger.overrun attribute.
11-84
Series 2600B System SourceMeter® Instrument Reference Manual
Affected by
Instrument reset
Recall setup
The type of measurement:
0 or display.MEASURE_DCAMPS: Current measurement function
1 or display.MEASURE_DCVOLTS: Voltage measurement function
2 or display.MEASURE_OHMS: Resistance measurement function
3 or display.MEASURE_WATTS: Power measurement function
Source-measure unit (SMU) channel (for example,
display.smua.measure.func applies to SMU channel A)
(on page 11-83)
Affected by
Where saved
Default value
Saved setup
1 (display.MEASURE_DCVOLTS)
Selects the current measure function
for SMU A.