Download Print this page

Tektronix KEITHLEY System SourceMeter 2636B Reference Manual page 528

Advertisement

Section 11: TSP command reference
meminfo()
This function returns the present amount of available memory and the total amount of memory in the instrument.
Type
TSP-Link accessible
Function
No
Usage
freeMem, totalMem = meminfo()
freeMem
totalMem
Details
This function returns two values:
The amount of free dynamically allocated memory available in kilobytes
The total amount of dynamically allocated memory on the instrument in kilobytes
The difference between the two values is the amount presently used.
Also see
None
node[N].execute()
This function starts test scripts from a remote node. This command is not available on the 2604B/2614B/2634B.
Type
TSP-Link accessible
Function
Yes (see Details)
Usage
node[N].execute("scriptCode")
N
scriptCode
Details
Only the remote master node can use the execute command to run a script on this node. This
function does not run test scripts on the master node; only on this node when initiated by the master
node.
This function may only be called when the group number of the node is different than the node of the
master.
This function does not wait for the script to finish execution.
This command should only be used from a remote master when controlling this instrument over a
®
TSP-Link
.
11-160
Series 2600B System SourceMeter® Instrument Reference Manual
Affected by
The amount of free dynamically allocated memory available
The total amount of dynamically allocated memory in the instrument
Affected by
The node number of this instrument
A string containing the source code
Where saved
Default value
Where saved
Default value
2600BS-901-01 Rev. E / January 2019

Advertisement

loading
Need help?

Need help?

Do you have a question about the KEITHLEY System SourceMeter 2636B and is the answer not in the manual?