Series 2600B System SourceMeter® Instrument Reference Manual
This function can also be used to start actions on the SMU in case of a missed trigger event.
Example
smua.trigger.arm.set()
Sets the arm event detector to the detected state for SMU channel A.
Also see
smuX.trigger.initiate()
smuX.trigger.measure.set()
smuX.trigger.source.set()
smuX.trigger.arm.stimulus
This attribute selects the event that will cause the arm event detector to enter the detected state.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = smuX.trigger.arm.stimulus
smuX.trigger.arm.stimulus = eventID
eventID
X
Details
Set this attribute to the event ID of any trigger event generator to wait for that event.
Set this attribute to zero to bypass waiting for events at the arm event detector (the SMU continues
uninterrupted through the remote trigger model). Set eventID to one of the existing trigger event IDs
shown in the following table.
2600BS-901-01 Rev. E / January 2019
(on page 11-262)
(on page 11-264)
(on page 11-273)
Affected by
SMU reset
Instrument reset
Recall setup
Event that triggers the arm detector
Source-measure unit (SMU) channel (for example,
smua.trigger.arm.stimulus applies to SMU channel A)
Section 11: TSP command reference
Where saved
Default value
Not saved
0
11-253