Series 2600B System SourceMeter® Instrument Reference Manual
localnode.passwordmode
This attribute stores the remote access password enable mode.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
mode = localnode.passwordmode
localnode.passwordmode = mode
mode
Details
This attribute controls if and where remote access passwords are required. Set this attribute to one of
the values below to enable password checking:
•
localnode.PASSWORD_NONE or 0: Disable passwords everywhere
•
localnode.PASSWORD_WEB or 1: Use passwords on the web interface only
•
localnode.PASSWORD_LAN or 2: Use passwords on the web interface and all LAN interfaces
•
localnode.PASSWORD_ALL or 3: Use passwords on the web interface and all remote command
interfaces
When using this command from a remote node, localnode should be replaced with the node
reference, for example node[5].passwordmode.
Example
mode = localnode.PASSWORD_WEB
localnode.passwordmode = mode
Also see
localnode.password
2600BS-901-01 Rev. E / January 2019
Affected by
Not applicable
The remote password enable mode
(on page 11-152)
Section 11: TSP command reference
Where saved
Default value
Nonvolatile memory
1 (localnode.PASSWORD_WEB)
Sets value of mode to PASSWORD_WEB.
Allows use of passwords on the web interface only.
11-153