Programs and program selection

The DAC Universal has 2 different programs that can be selected as
needed. They are set using the menu system, see "Menu system.
[ → 65]".
Factory default is the standard program (134 °C / 273 °F, sterilization time
3 minutes).
Steri
Total
Program
lization
cycle
time
duration*
3 min
18 min/
Standard program
21 min
18 min
33 min
Prion program (required in
France, for example)
* with factory settings
IMPORTANT
The DAC Universal remembers the last program used, even after being
switched off. This also happens in case the electric power supply to the
DAC Universal was interrupted.
IMPORTANT
Make sure the correct lid type is selected.
It is possible to block programs that are not used. Please contact your
authorized dealer to block/unblock programs.