2 About the Software
A
NALYSIS
Use to perform an analysis. When analyzing a sample, ensure the insert and cup in the selected
method match the insert and cup being used.
Some samples require external preparation prior to placing in the sample chamber. This can
remove volatiles from the sample that would otherwise damage internal components or impact the
pressure readings. When moisture is present on a sample, the vapor pressure impacts the
pressure readings.
Maximum accuracy is obtained when the sample chamber is nearly filled with sample. The
minimum suggested sample quantity is 10% of the available volume within the sample chamber. If
a small sample quantity is available, inserts can be used to further reduce the available sample
chamber volume.
1. Tap Analysis.
Settings >
ator fields are shown.
2. Select a Method from the list. Standard is the default unless another method has been
configured and enabled as the default. To add a new method, refer to Settings >
on page 2 -
21.
3. Enter a Description to display when results are shown on screen, printed, or exported.
4. Enter text to display that identifies the Operator. This field is only available if the Operator
function has been enabled under Settings >
5. Enter the sample Mass.
6. Tap Next.
2 - 8
Display on page 2 - 40
determines if the Method, Description and Oper-
Display on page 2 -
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
Method
40.