Unable to connect to Wi-Fi.
Cause A: The password is incorrect.
Action A: Contact your IT department and ensure the password is correct.
Cause B: The Wi-Fi network is inactive or has low signal strength near the instrument.
Action B: Contact your IT department and ensure the network is active and has adequate
signal strength near the instrument.
Cause C: The instrument does not have appropriate permissions.
Action C: Contact your IT department to change instrument permissions.
Cause D: An internal processing and/or hardware error has occurred.
Action D: Contact your Micromeritics Service Representative.
Unable to mount network drive.
Cause A: The entered parameters are incorrect.
Action A: Contact your IT department and ensure all parameters are correct.
Cause B: Network drive is not connected to the network.
Action B: Contact your IT department and ensure you can connect to the network drive
from other computers on the same network.
Cause C: Instrument is not connected to the network. Or, it is connected to a different
network than the network drive.
Action C: Contact your IT department. If the instrument is on the network, you should be
able to connect to it remotely as described in the operator manual.
Cause D: The instrument does not have appropriate permissions.
Action D: Contact your IT department to change instrument permissions.
Cause E: An internal processing and/or hardware error has occurred.
Action E: Contact your Micromeritics Service Representative.
Warning: Could only achieve pressure of [n] when trying to reach [n].
Cause: The operation could only fill to 95% of the specified gauge pressure. The
operation is allowed to continue. If desired, perform the actions below.
Action A: Increase the regulator pressure, or, if there is insufficient gas in the tank, obtain a
new tank. Tank pressure should be at least 200 psi above regulator pressure.
Action B: Check the instrument for gas leaks following the instructions in the operator
manual.
Action C: Contact your Micromeritics Service Representative if this error message con-
tinues.
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
C Error Messages
C - 7
Need help?
Do you have a question about the ACCUPYC III and is the answer not in the manual?