Gases
Research grade helium is recommended. If unavailable, use helium with a dew point of -67 °C
(-88 °F) or lower.
Sample Cups
3
10 cm
chamber
3
100 cm
chamber
Analysis
Precision
Accuracy
Gas Consumption
Temperature Control
Range
Stability
Due to continuous improvements, specifications are subject to change without notice.
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
1.85 cm ID × 3.95 cm D (0.72 in. ID × 1.55 in. D)
4.62 cm ID × 6.18 cm D (1.82 in. ID × 2.43 in. D)
Reproducibility is typically to within ±0.01% of the nominal full-
scale sample chamber volume. Reproducibility is guaranteed to
within 0.015% of the nominal full-scale volume on clean, dry,
thermally equilibrated samples using helium.
Accurate to within 0.02% of (nominal full-scale sample chamber
volume + measured sample volume).
Approximately 1 cm³ STP times the nominal cell volume per ana-
lysis cycle.
4 to 60 °C for ATC units
±0.025 °C
Specifications for AccuPyc III