Micromeritics ACCUPYC III Operator's Manual page 66

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC III:
Table of Contents

Advertisement

2 About the Software
H
ISTORY
Displays all calibrations performed and includes date, sample and reference-chamber volumes. In
addition to the chamber calibration values, a Result column records whether the calibration was
"Applied," "Rejected," "Factory," or the result of a "Reset."
Tap
Export
and select an export type (such as PDF or TXT).
C
ALIBRATION
Performs a chamber volume calibration. The first screen provides information about the action to
be performed.
Good lab practice is to perform a verification on a regular schedule. 90 days is a
reasonable period, or when major changes are made to normal operating conditions
(such as temperature differences or switching gases or if a change to the internal
chamber is suspected).
If a verification passes, there is no need to calibrate. If a verification fails, a calibration
should be done and the cup volume remeasured.
This process can take several hours. During the process a message displays
indicating that the sample chamber should be empty. Tap Next. As the process
continues, a message displays indicating to insert the calibration standard. Tap
to continue the process.
2 - 36
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023