Exterior surfaces should be cleaned with soap or mild detergents. Cleaning agents should be
applied with a soft cloth. Do not use solvents or abrasives as they may damage some surfaces.
The air intake filters at the rear of the instrument should be inspected periodically and rinsed or
replaced if they are clogged. Clogged filters may impair the instrument's temperature control and
could damage the TEC.
A fan is used to exchange heat with the TEC module. Air is pulled from the underside
at the back of the instrument and exhausted out the back. Ensure that the back of the
instrument is not obstructed by cables or debris. Ensure that area behind the fan grille
is clear. Obstructed airflow could impair temperature control and could damage the
TEC.
Do not immerse the pycnometer or the power cord in any liquids. Doing so could res-
ult in electrical shock to personnel or damage to the unit.
Do not allow liquid to penetrate the casing of the pycnometer. Doing so could result in
damage to the unit.
D
ECONTAMINATION OF THE
Always wear personal protection equipment appropriate for the type and level of con-
tamination.
In addition to following the instructions to clean the pycnometer, service personnel should use
compressed air to decontaminate internal components.