S
C
YSTEM
OMPONENTS
V
S
ERIFICATION
TANDARD
Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n
Do not drop the standard into the sample cup. Gently place the standard into the cup to prevent
n
damage.
Always return the standards to the storage location inside the front panel of the instrument.
n
Standards are unit specific.
n
S
C
AMPLE
HAMBER
Wear latex or nitrile, powder-free gloves to prevent transfer of oil from hands.
n
Keep the sample chamber cap closed except when inserting or removing the sample cup.
n
When left open, the sample chamber temperature stability may be affected
and/or water vapor will adsorb on the inner surface of the chamber. Either of
these conditions can affect analysis results.
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
C
AP
System Components
1 - 5