1 About the AccuPyc III
T
C
EMPERATURE
ONTROL
Density measurements at specific temperatures are required for certain applications. The
AccuPyc III TEC maintains an accurate temperature control during analysis without external
supplies or systems.
All units use a thermoelectric cooler (TEC) for stability. ATC units allow setting the chamber
temperature from 4 °C to 60 °C.
Adequate ventilation is required to ensure effective temperature control and system
longevity. Ensure that the underside and back of the instrument are free of obstruction
and receive sufficient air flow.
The TEC air intake may gather dust and impede performance. The intake filter should
be inspected weekly and cleaned with a duster or compressed air, if necessary. If
extremely dirty, wash and dry overnight, especially if in a dirty environment.
1 - 6
(TEC)
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023