Micromeritics ACCUPYC III Operator's Manual page 103

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC III:
Table of Contents

Advertisement

Action D:
Contact your Micromeritics Service Representative.
Last measured atmospheric pressure is [n]. Possible vent valve failure or vent
blockage.
Cause: Vent valve failure or Vent Port blockage. This can manifest as a high atmospheric
pressure measurement, which causes a failure to fill to the requested gauge pres-
sure.
Action A: Ensure the Vent Port on the back of the unit is clear of dust and other obstruc-
tions.
Action B:  Contact your Micromeritics Service Representative if this error message con-
tinues.
Maximum evacuation time ([n] s) exceeded. [n] canceled.
Cause: The target vacuum pressure could not be reached during vacuum preparation.
Action A: Ensure the vacuum pump is attached and running.
Action B:  Check the instrument for gas leaks following the instructions in the operator
manual.
Action C: Contact your Micromeritics Service Representative if this error message con-
tinues.
Maximum fill time ([n] s) exceeded. [n] canceled.
Cause: There was insufficient pressure to allow filling.
Action A: Increase the regulator pressure, or, if there is insufficient gas in the tank, obtain a
new tank. Tank pressure should be at least 200 psi above regulator pressure.
Action B:  Check the instrument for gas leaks following the instructions in the operator
manual.
Action C: Contact your Micromeritics Service Representative if this error message
continues.
No legacy instrument at the address.
Cause A: The IP address is incorrect.
Action A: Ensure the IP address matches the IP address on the legacy instrument's
screen.
Cause B:  Legacy instrument is not connected to the network (if you are connecting both
instruments to the lab network).
Action B: Contact your IT department and ensure you can connect to the legacy
instrument from other computers on the same network.
Cause C: Instrument is not connected to the network (if you are connecting both
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
C Error Messages
C - 3