G
R
AS
EQUIREMENTS
Guidelines for Connecting Gases on page 3 - 59
Improper handling, disposing of, or transporting potentially hazardous materials can
cause serious bodily harm or damage to the instrument. Always refer to the SDS
when handling hazardous materials. Safe operation and handling of the instrument,
supplies, and accessories are the responsibility of the operator.
Toxic, corrosive, flammable, poisonous, or injurious gases should not be used with
the AccuPyc.
The pycnometer uses helium, nitrogen, or other dry, inert gas (99.995% pure or better) to provide
rapid, accurate analyses. It is recommended that the cylinder containing the analysis gas be fitted
with a gas regulator set for 22 psig (152 kPag). The pressure input to the pycnometer should
never be greater than this pressure.
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
Gas Requirements
1 - 15
Need help?
Do you have a question about the ACCUPYC III and is the answer not in the manual?
Questions and answers