C Error Messages
instruments to the lab network). Or, it is connected to a different network than the
legacy instrument.
Action C: Contact your IT department. If the instrument is on the network, you should be
able to connect to it remotely as described in the operator manual.
Cause D: The instrument is not directly connected to the legacy instrument.
Action D: Connect them by plugging an Ethernet cable into the appropriate ports as
described in the operator manual. Cycle power on both instruments.
Cause E: An internal processing and/or hardware error has occurred.
Action E: Contact your Micromeritics Service Representative.
No printer available.
Cause A: No printer is connected to the network.
Action A: Contact your IT department and ensure you can print to at least one printer from
other computers on the same network.
Cause B: Instrument is not connected to the network. Or, it is connected to a different
network than the printer.
Action B: Contact your IT department. If the instrument is on the network, you should be
able to connect to it remotely as described in the operator manual.
Cause C: The instrument does not have appropriate permissions.
Action C: Contact your IT department to change instrument permissions.
Cause D: An internal processing and/or hardware error has occurred.
Action D: Contact your Micromeritics Service Representative.
Other operations did not finish after [n] s. [n] canceled.
Cause: Background operations did not finish, so the desired operation cannot be started.
Action: Cycle power. Contact your Micromeritics Service Representative if this error mes-
sage continues.
Pressure overrange [n] detected. [n] canceled.
Cause: Pressure reading is above normal operating range, so the operation was can-
celed.
Action A: Adjust the regulator pressure so that it is no greater than 22 psig (152 kPag).
Action B:  Contact your Micromeritics Service Representative if this error message con-
tinues.
C - 4
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023