Safe Servicing - Micromeritics ACCUPYC III Operator's Manual

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC III:
Table of Contents

Advertisement

3 Maintenance and Troubleshooting
S
S
AFE
ERVICING
Do not service or modify this instrument without the authorization of Micromeritics
Service Personnel.
To ensure safe servicing and continued safety of the instrument after servicing, service personnel
should be aware of the following risks:
Product specific risks that may affect service personnel:
Electrical. Servicing or repair could require opening the outer panels and exposing ener-
n
gized electrical components.
High temperatures. Temperature controlled components internal to the instrument may be
n
hot and could pose a burn hazard to service personnel.
Protective measures for these risks:
Electrical. The electrical components operate at low voltage (24V or less) and pose low risk
n
when energized. However, maintenance, troubleshooting, and repairs should be performed
with the instrument de-energized whenever possible, in accordance with standard electrical
safety guidelines.
High temperatures Ensure the temperature control is off and verify temperature control
n
components are near ambient temperature before servicing.
Verification of the safe state of the instrument after repair:
All instrument panels and covers installed.
n
Gas lines connected and pressurized to normal operating pressure with no leaks.
n
3 - 52
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023