Block temperature is unresponsive to cooling after applying current below [n] A for [n]
s. Starting temp.: [n] °C, current temp.: [n] °C.
Block temperature is unresponsive to heating after applying current above [n] A for [n]
s. Starting temp.: [n] °C, current temp.: [n] °C.
[n] temperature too high, above [n] °C.
Cause:
An internal processing and/or hardware error has occurred. The instrument
automatically turns off the TEC as a safety precaution.
Action:
Cycle power. Contact your Micromeritics Service Representative if this error mes-
sage continues.
Cannot actuate hardware. No communications with the PCB.
Cannot communicate with the instrument hardware.
Cannot connect to instrument hardware.
Cannot get communications attributes.
Cannot get the PCB Status.
Cannot install software file: [n]
Cannot reach server [n]: [n]
Cannot save copy.
Cannot set communications attributes.
Cannot start an operation. No communications with the PCB.
Could not fetch results.
Error generating PDF file for report.
Error generating plot for PDF report.
Error logger cannot be initialized. Error code [n]. Program will terminate.
Hardware error. Unable to set temperature.
Hardware PCB Error: [n]
Cause: An internal processing and/or hardware error has occurred.
Action: Cycle power. Contact your Micromeritics Service Representative if this error mes-
sage continues.
Cannot export to USB. Multiple USB drives found.
Cause: Multiple USB drives are plugged in.
Action: Unplug the other USB drives.
Cannot export to USB. No USB drive found.
Cause: USB drive is not plugged in.
Action: Plug in USB drive.