Inserts - Micromeritics ACCUPYC III Operator's Manual

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC III:
Table of Contents

Advertisement

I
NSERTS
Settings > Inserts
Use to configure the type of insert being placed in the chamber: Multivolume Insert, Volume
Standard, Cup and Measure Volume.
Inserts are used to reduce the available volume within the sample chamber. This allows maximum
accuracy to be obtained when using lower sample quantities. Inserts are aluminum cylinders that
act as a holder for a new, lower-volume sample cup.
When measuring the volume of a cup, the reference object that came with the instrument or
multivolume kit should be used. When measuring a multivolume insert, only the insert needs to be
placed in the chamber.
After the volume has been measured, the software will create a file in which the name can be
modified. The measured volume and temperature at which the volume was obtained cannot be
modified.
Changing these settings doesn't configure the instrument for a particular insert or cup.
After an insert or cup is added, it can be selected for a particular method by editing the
method's settings (refer to Settings >
Also, if an attempt is made to delete an insert or cup linked to an analysis or
calibration, an error message displays indicating the insert or cup is still referenced.
To add a new Multivolume Insert, Volume Standard, or Cup, tap the
applicable details. (Measure Volume has a different set of fields than those shown below.)
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
Method on page 2 -
21).
Plus

Inserts

sign and enter the
2 - 31