C
HAMBER
Settings > Chamber
Used to verify proper calibration of the instrument (including when the last verification ran and the
verification history) and run supplemental instrument operations.
The data collected for Last and History can be printed or exported.
V
ERIFY
This process can take several hours. During the process, a prompt displays that
indicates to insert the calibration standard. Tap
1. If changes are required to the information already displayed, tap in the Volume Standard or
Fill Pressure fields and enter the applicable values. Nominal cell volume is either 10 cm
3
100 cm
, depending on the AccuPyc III size in use.
2. Place the selected volume standard in the sample chamber.
3. Tap Next. The verification process takes place. The measured volume should be within
tolerance of the volume standard. The tolerance is given as:
Tolerance = (nominal cell volume * 0.02%) + (standard volume * 0.02%)
4. Once completed, the other tabs on this screen update.
5. To stop the verification, tap Cancel.
AccuPyc III Operator Manual
135-42800-01 (Rev -) — Nov 2023
Next
to continue the process.
Chamber
3
or
2 - 27