Display Format
MEAS DISPLAY Page

Monitor Information

The following describes monitor information displayed on the MEAS DISPLAY
page. This information cannot be changed on the MEAS DISPLAY page.
Monitor
information
VAC
IAC
VDC
IDC
CORR
CH
1. This can be displayed only when option 001 is installed.
2. SINGLE is displayed when option 301 is not installed.
The VDC/IDC monitor values are the ones measured under the condition
that the setting value of the test signal level (LEVEL field) is output.
Therefore, these values are different from the measurement results
displayed when the measurement parameter is DC voltage
measurement/DC current measurement (Vdc-Idc).
Keysight E4980A/AL Precision LCR Meter
Description
Displays the test signal voltage level (AC voltage). Displays
measurement results regardless of the type of measurement
parameter
Displays the test signal current level (AC current). Displays
measurement results regardless of the type of measurement
parameter
Displays the DC bias voltage level (DC voltage). Set the VDC MON
field of the [Means Setup] page to ON when you want to display it.
Displays the DC bias current level (DC current). Set the IDC MON field
of the [Means Setup] page to ON when you want to display it
Displays the correction information of enabled correction types
(OPEN/SHORT/LOAD) and cable length
Displays the channel number in the multi-correction mode
1
2
75