Open Correction

The open correction feature of the E4980A/AL compensates for any stray
admittances (G, B) that may exist within the interval from the calibration plane,
which is determined by the selected cable length, to the DUT connecting
points (see
Figure 4-9
Stray admittances
Description of Correction Based on All Frequency Points
In this correction method, open correction data are collected from all 51
predetermined frequency points, regardless of the test frequency settings. At
the other frequency points, correction is done using data calculated by
interpolating the data collected at the 51 points (see
following tables list the 51 predetermined frequency points:
Table 4-7
Predetermined frequency points in correction based on all frequency points
Test frequency points
20 Hz
25 Hz
30 Hz
40 Hz
50 Hz
60 Hz
80 Hz
Keysight E4980A/AL Precision LCR Meter
Figure
4-9).
100 Hz
1 kHz
120 Hz
1.2 kHz
150 Hz
1.5 kHz
200 Hz
2 kHz
250 Hz
2.5 kHz
300 Hz
3 kHz
400 Hz
4 kHz
500 Hz
5 kHz
600 Hz
6 kHz
800 Hz
8 kHz
Figure
4-10). The
10 kHz
100 kHz
12 kHz
120 kHz
15 kHz
150 kHz
20 kHz
200 kHz
25 kHz
250 kHz
30 kHz
300 kHz
40 kHz
400 kHz
50 kHz
500 kHz
60 kHz
600 kHz
80 kHz
800 kHz
1 MHz
1.2 MHz
1.5 MHz
2 MHz