List:bias:current; List:bias:voltage; List:clear:all - Keysight E4980AL User Manual

Precision lcr meter
Hide thumbs Also See for E4980AL:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
[Meas Setup] - LIST SETUP - LOW
[Meas Setup] - LIST SETUP - HIGH

:LIST:BIAS:CURRent

Syntax
:LIST:BIAS:CURRent <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:CURRent?
Description
Clears the previous list sweep table, sets DC current sweep as the list sweep
parameter, and sets DC current points for the DC current list sweep. If the list
sweep is not set for DC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Range of n
Unit
Equivalent key
[Meas Setup] - LIST SETUP - BIAS[A]

:LIST:BIAS:VOLTage

Syntax
:LIST:BIAS:VOLTage <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:VOLTage?
Description
Clears the previous list sweep table, sets DC voltage sweep as the list sweep
parameter, and sets DC voltage points for the DC voltage list sweep. If the list
sweep is not set for DC voltage, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Range of n
Unit
Equivalent key
[Meas Setup] - LIST SETUP - BIAS[V]

:LIST:CLEar:ALL

Syntax
:LIST:CLEar:ALL
358
<point n>
1 to 201
A
<point n>
1 to 201
V
Keysight E4980A/AL Precision LCR Meter