SCPI Command Reference
E4980A/AL commands
Equivalent key
No equivalent key is available on the front panel.
*WAI
Syntax
*WAI
Description
Waits until all commands sent before this command have been executed. The
following commands are then executed. (No Query)
Equivalent key
No equivalent key is available on the front panel.
:ABORt
Syntax
:ABORt
Description
Resets the trigger system. The trigger state is then in the "Idle" state on the
state diagram. (No Query)
Equivalent key
[Meas Setup] - MEAS SETUP - REF A - MEAS ABORT
[Meas Setup] - MEAS SETUP - B - MEAS ABORT
[Meas Setup] - CORRECTION - OPEN - MEAS OPEN - ABORT
[Meas Setup] - CORRECTION - SHORT - MEAS SHORT - ABORT
:AMPLitude:ALC
Syntax
:AMPLitude:ALC {ON|OFF|1|0}
:AMPLitude:ALC?
Description
Enables the Automatic Level Control (ALC).
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - MEAS SETUP - ALC
:APERture
Syntax
:APERture {SHORt|MEDium|LONG},<numeric>
:APERture?
Description
Sets the measurement time mode and the averaging rate.
Keysight E4980A/AL Precision LCR Meter
Description
Enables the Automatic Level Control
Disables the Automatic Level Control
325