High Vt E Alarm (High Expired Tidal Volume) - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.3.6
High Vt
Property
Alarm text
Priority
Alarm condition
Possible cause
Setting range
Setting resolution
Ventilator action
Doc. 006150 M-1
Alarm (High Expired Tidal Volume)
e
Description
High Vte
Medium
A High Expired Tidal Volume alarm will be given when the
monitored Expired Tidal Volume exceeds the alarm limit for
15 seconds.
This alarm is only used if the Vivo 65's patient circuit type is set
to Dual limb or Leakage (single limb) circuit.
Mismatch between Expired Tidal Volume and alarm setting.
Mismatch between selected and used patient circuit.
Pressure settings causing the Expired Tidal Volume to exceed
the set alarm level.
300 ml (adult), 50 ml (paediatric) to 3000 ml (adult), 600 ml
(paediatric)
Off
10 below 600 ml, 100 above 600 ml
The ventilator will continue treatment with the same settings.
Alarms
95
User manual