Using The Vivo 65 With The Fio 2 Sensor - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

1 Connect the nurse call cable on the left
side panel of the Vivo 65.
2 Test the connection by triggering an alarm on the Vivo 65 and verify that the nurse call
system activates.
5.8.3
Using the Vivo 65 with the FiO
The FiO
urements. The FiO
Vivo 65. The FiO
loaded to a PC and viewed in the Vivo 65 PC software.
The FiO
The FiO
NOTE
FiO
calibration can be performed by a clinician.
2
CAUTION!
Note that the operating conditions for the FiO
system conditions. If the sensor is used outside its operating conditions the FiO
urements might deviate.
NOTE
The FiO
pressure.
Doc. 006150 M-1
sensor can be connected to the Vivo 65 in order to monitor and store FiO
2
sensor measures the fraction of inspired oxygen at the air outlet of the
2
measurements will be stored in the data memory which can be down-
2
sensor is not intended to be in contact with the patient's body.
2
sensor should be calibrated when first connected and then at least once a month.
2
monitoring automatically compensates for changes in ambient barometric
2
Sensor
2
sensor is different from the ventilator
2
meas-
2
meas-
2
How to Use the Vivo 65
User manual
79