Apnea Alarm; Disconnection Alarm - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.3.15

Apnea Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
6.3.16

Disconnection Alarm

CAUTION!
No single alarm can reliably detect all disconnections due to the number of possible
combinations of therapy settings, circuit configurations and patient interfaces. To verify
that patient disconnection can be detected, including if the patient interface becomes
accidentally detached from the patient, it is advised to test the functionality of the Dis-
connection Alarm upfront with the complete set-up as used during treatment, including
items such as filters, circuit, connectors, interface (mask, cannula etc.)
104
Alarms
User manual
Description
Apnea
High
An Apnea alarm will be given when no patient-triggered breath
is detected for the set period of time.
Patient stopped breathing.
Patient decreases spontaneous breathing.
Circuit disconnection.
Inspiratory Trigger is set too high.
Inspiratory effort detected by the Vivo 65.
The Vivo 65 will continue treatment according to the current
settings.
5 to 60 s. (Non MPV mode)
15 to 900 s. (MPV mode)
Off
5 s below 15 s.
15 s above 15 s.
MPV mode:
15 s below 60 s.
60 s above 60 s.
Doc. 006150 M-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivo 55

Table of Contents