6.4.24

Alarm for Beeper Failure

Property
Alarm text
Alarm condition
Priority
Ventilator action
6.4.25

Int. Function Failure (Internal Function Failure Alarm)

Property
Alarm text
Alarm condition
Priority
Ventilator action
Ventilator reset
6.5

Alarm Test

6.5.1

Instructions

NOTE
To perform the alarm test, follow the instructions below:
Low Pressure and Disconnection Alarms
1 Start treatment and disconnect the patient circuit.
2 Wait 15 seconds.
3 The Low Pressure Alarm and/or the Disconnection Alarm will be given.
4 Stop treatment. Test completed.
Low Vti or Low Vte Alarms
1 Start treatment and block the patient circuit completely to simulate an occlusion.
2 Wait 15 seconds.
122
Alarms
User manual
Description
Alarm Beeper Fail
Failure of the beeper control by the treatment processor.
High
The ventilator will continue treatment with the same settings.
Description
Int. function failure:
An Internal Function Failure alarm will be given when the ven-
tilator has an internal function failure, followed by an error
code for the specific failure. All Internal Function Failure alarm
error codes are defined and explained in the ventilator Service
Manual.
High
The ventilator will stop the treatment and shut down.
Restart the ventilator.
In order to test an alarm it must be activated by your clinician.
In some cases a setting must be altered.
Doc. 006150 M-1