Advertisement

MONITORED
VALUE
% Spont
SpO
2
Pulse Rate
I:E
Insp. Time
Rise Time
EtCO
InspCO
* BTPS (Body Temperature and Pressure Saturated)
** Effects of humidity on FiO
8.2.3

Alarms

Please refer to the information in 6 Alarms, page 88.
8.2.4

Technical Specifications and Conditions

8.2.4.1 Power Supply
POWER SUPPLY
AC Power (Mains) supply
External battery
Click-on battery*
Internal battery
*For transportation by air, please observe that the click-on battery capacity is 192 Wh which
exceeds stated limits. Always consult your airline company for transportation restrictions.
Doc. 006150 M-1
RANGE
0 to 100%.
70 to 100%.
18 to 300 bpm.
1:99 to 10:1.
0.3 to 5 s.
0.1 to 5 s.
0 to 25%.
2
0 to 25%.
2
accuracy: -0.03% per % RH relative to calibration point
2
SPECIFICATION
100 to 240 V AC, tolerance: +10%/-20%, 50 to 60
Hz, max 300 VA.
24 V DC, tolerance: 24 V ± 6 V.Max 7 A, 140 W.
Capacity 5.2 Ah. LiIon. Operational time 8 hours, life-
time 3 years.
Capacity 2.6 Ah. LiIon. Operational time 4 hours, life-
time 3 years.
ACCURACY
Not applicable
±3 digits. Data update period 1
s.
4-beat average signal processing.
±3 digits. Data update period 1
s.
4-beat average signal processing.
±0.1 unit
±0.1 s
±10% or ±0.1 s, whichever is
greatest
0 to 15%: ±(0.2 vol% + 2% of
reading), 15 to 25%: unspecified
0 to 15%: ±(0.2 vol% + 2% of
reading), 15 to 25%: unspecified