3 The Low Vti or Low Vte Alarm will be given.
4 Stop treatment. Test completed.
Obstruction Alarm
1 Start treatment; block the patient circuit completely to simulate an obstruction.
2 Wait approximately 10 seconds.
3 The Obstruction Alarm will be given. (If the expiratory path is blocked, a
Disconnection Alarm may be triggered.)
4 Stop treatment. Test completed.
Alarms
123
User manual
Doc. 006150 M-1