Low Pulse Rate Alarm; Technical Alarms; Power Failure Alarm - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.3.27

Low Pulse Rate Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO
6.4

Technical Alarms

6.4.1

Power Failure Alarm

Property
Alarm text
Priority
Alarm condition
Reset criteria
Ventilator action
112
Alarms
User manual
Description
Low Pulse Rate
High
A low pulse rate alarm will be given when the measured pulse
rate goes below the alarm limit for 15 seconds.
Bad positioning of the finger probe.
Too low flow of bleed-in oxygen.
Insufficient ventilatory support.
The pulse rate goes back above the alarm limit.
The Vivo 65 will continue treatment according to the current
settings.
30 to 230 bpm (beats per minute)
Off
5 bpm (beats per minute)
sensor.
2
Description
No alarm text during the alarm.
After restart: Treatment was restarted after power failure.
High
A Power Failure alarm will be given when the last power source
is below limits.
The power is restored
The Vivo 65 stops treatment and gives alarm for at least
2 minutes.
If power is restored within 10 minutes, the Vivo 65 will auto-
matically resume treatment with current settings.
Doc. 006150 M-1