Signal); Alarm); Disconnected; Co2 Accuracy Unspecified - Breas VIVO 65 User Manual

Hide thumbs Also See for VIVO 65:
Table of Contents

Advertisement

6.4.14
SpO
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
6.4.15
CO
2
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
6.4.16
CO
2
Property
Alarm text
Alarm condition
Priority
Ventilator action
118
Alarms
User manual
Signal Inadequacy Alarm (SpO
2
Description
SPO2 Signal
An SpO
probe is not able to perform an adequate measurement, due to
low perfusion or artefacts.
Check the SpO
High
The ventilator will continue treatment with the same settings.
Disconnected (CO
Description
CO2 Sensor Disconnected
A CO
communication between the ventilator and the CO
been lost for 2 seconds. Check the CO
High
The ventilator will continue treatment with the same settings.
Confirmation by user or reconnected/changed.
Sensor Accuracy Unspecified Alarm (CO
Description
CO2 Accuracy Unspec.
A CO
an unspecified accuracy in the CO
Perform a zeroing procedure of the CO
High
The ventilator will continue treatment with the same settings.

Signal)

2
Signal Inadequacy alarm will be given when the SpO
2
sensor.
2
Bad positioning or occlusion of the probe.
Low blood flow in finger.
Sensor Failure/Disconnection Alarm)
2
Sensor Failure/Disconnection alarm will be given when
2
CO
Sensor disconnected.
2
Failure in the CO
sensor.
2
Sensor Accuracy Unspecified alarm will be given when
2
sensor.
2
Accuracy Unspec)
2
measurement has occurred.
2
sensor.
2
2
sensor has
2
Doc. 006150 M-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivo 55

Table of Contents