6.4.14
SpO
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
6.4.15
CO
2
Property
Alarm text
Alarm condition
Priority
Possible cause
Ventilator action
Reset
6.4.16
CO
2
Property
Alarm text
Alarm condition
Priority
Ventilator action
118
Alarms
User manual
Signal Inadequacy Alarm (SpO
2
Description
SPO2 Signal
An SpO
probe is not able to perform an adequate measurement, due to
low perfusion or artefacts.
Check the SpO
High
•
•
The ventilator will continue treatment with the same settings.
Disconnected (CO
Description
CO2 Sensor Disconnected
A CO
communication between the ventilator and the CO
been lost for 2 seconds. Check the CO
High
•
•
The ventilator will continue treatment with the same settings.
Confirmation by user or reconnected/changed.
Sensor Accuracy Unspecified Alarm (CO
Description
CO2 Accuracy Unspec.
A CO
an unspecified accuracy in the CO
Perform a zeroing procedure of the CO
High
The ventilator will continue treatment with the same settings.
Signal)
2
Signal Inadequacy alarm will be given when the SpO
2
sensor.
2
Bad positioning or occlusion of the probe.
Low blood flow in finger.
Sensor Failure/Disconnection Alarm)
2
Sensor Failure/Disconnection alarm will be given when
2
CO
Sensor disconnected.
2
Failure in the CO
sensor.
2
Sensor Accuracy Unspecified alarm will be given when
2
sensor.
2
Accuracy Unspec)
2
measurement has occurred.
2
sensor.
2
2
sensor has
2
Doc. 006150 M-1