Volume Control Modes
The worst case Vivo 65 System is the dual limb patient circuit with or without HCH
humidifier, bacterial filter, FiO
8.2
Data Parameters
NOTE
All stated tolerances includes measurement uncertainty. The accuracies have been tested
with all allowed configurations. Stated tolerances only discloses the maximum tolerance.
If a parameter's tolerance is described with both absolute and relative measures, the
greater one applies if nothing else is stated. Percent measurements are relative to the set
value.
8.2.1
Settings
Please refer to the information in 5.4 Functions and Parameters in the Vivo 65, page 56.
8.2.2
Monitored Values
MONITORED
VALUE
P
peak
PEEP
P
mean
Leakage
MV
MV
Vt
i
Vt
e
FiO
% in TgV
Total Rate
Spont Rate
134
Technical Specifications
User manual
RANGE
4 to 70 cmH
0 to 30 cmH
0 to 70 cmH
0 to 100 l/min (BTPS*).
0 to 99.9 l (BTPS*).
i
0 to 99.9 l (BTPS*).
e
0 to 9999 ml (BTPS*).
0 to 9999 ml (BTPS*).
0 to 100%.
2
0 to 100%.
0 to 99 bpm.
0 to 99 bpm.
sensor and CO
sensor.
2
2
O.
2
O.
2
O.
2
ACCURACY
±0.5 cmH
O or ±10%, which-
2
ever is greatest
±0.5 cmH
O or ±10%, which-
2
ever is greatest
±0.5 cmH
O or ±10%, which-
2
ever is greatest
±10%
±10% or (±10 ml × bpm),
whichever is greatest
±10% or (±10 ml × bpm),
whichever is greatest
±10 ml or 10%, whichever is
greatest
±10 ml or 10%, whichever is
greatest
±2%**
±1%
1 bpm
1 bpm
Doc. 006150 M-1