CAUTION!
36
Preparing the Vivo 65 for Use
User manual
To optimize the treatment with Vivo 65's paediatric settings in Paediatric mode
when using a dual limb circuit, always use a paediatric dual limb insert and a
15 mm dual limb circuit. Using a paediatric dual limb insert with an adult (22 mm)
dual limb circuit may impair the performance of the ventilation treatment and
deteriorate the measurement precision of the exhaled air.
Always perform a pre-use test after changing the patient circuit type and/or insert.
An incorrect combination of patient circuit type and insert will result in a Circuit
Type/Insert Mismatch alarm, with medium alarm priority.
An incorrect combination of patient mode and insert will result in a Patient Mode/
Insert Mismatch alarm, with medium alarm priority.
Do not remove the gasket that is placed inside the Vivo 65's insert socket. Remov-
ing the gasket will result in a patient circuit leakage and a Disconnection alarm
and/or Exhalation Valve Control Error alarm.
Doc. 006150 M-1