9 Trouble-shooting
Device status and
Device status message
diagnosis message
(Display)
(Control system)
Zeropoint adjustment
P
$
is not possible
No communication to
–
amplifier
134
No.
0-ADJ. FAIL CH2
744
# 744
–
PROline Prosonic Flow 93 PROFIBUS-DP/-PA
Initial status
Measured
Analog Input Block/
value Quality /
Totalizer Block
Substatus /
Alarm limit
non specific
UNCERTAIN
0x40
no limits
device failure
BAD
0x0F
constant
Cause/remedy
Cause:
Zero point adjustment of channel 2
is not possible or was cancelled.
Remedy:
Make sure that zero point
adjustment is carried out at "zero
o
flow" only (v = 0 m/s)
Page 113.
Cause:
Communication error. No
communication to the amplifier.
Remedy:
1. Switch power supply off and
then on again
2. Check whether the electronics
boards are correctly inserted in
o
their holders
Page 138