Advertisement

9
Possible causes and remedies
The device is switched on, but is not reachable due to erroneous parameteri-
zation of the communication parameters.
Ensure that the communication parameters in the
Modbus AFB
are coordinated between the master and slave and/or client and
server.
OR
[only Slave/Server]
The error tolerance during data import or data export of a register is too low.
On a test basis, increase the Error filter parameter under:
Modbus –
Communication mode: Master or client –
<Remote device x>, tab Parameter, section General.
OR
[Master and slave]
At least one register is incorrectly or not at all parameterized on one of the two
sides.
Use the operation panel of the enCore device in the display belonging to the
Modbus AFB
and check which registers are not regularly refreshed. On both
sides, check for errors in the respective parameterization concerning these
register numbers.
OR
[Master and slave]
The cabling is defective or there is a cable breakage.
Check the cabling and replace the damaged cable.
OR
[only master]
The remote device (slave or server) processes the telegrams slower than expec-
ted.
On a test basis, increase the Transaction timeout of the master under:
Modbus –
Communication mode: Master or client, tab Parameter, section
Interface

9.7 Sensor errors

Sensor errors, usually recognizable by the values ​ ​ d isplayed in red on
play Q2 sensor values
block.
Inform Honeywell
analyzers
GasLab Q2
(p. 146), can only be be corrected by replacing the sensor
12.1 Technical support Flow Computers and Gas
➪  
(p. 293)
User Manual (N )
Possible malfunctions and troubleshooting
Basic System and
Remote devices –
7.3.10 Dis-
➪  
258 of 315