To take the enCore device out of operation, proceed as follows:
1. Disconnect the device from the power supply.
2. Turn off all gas flows and depressurize all gas inlets.
3. Here, too, prevent the unintentional opening of the shut-off valves by taking suit-
able measures so that no danger can arise from further disassembly.
4. In a non-explosive atmosphere, remove all connected gas lines, cables and data
lines.
5. Dismantle the device mechanically.
Next steps:
In case of temporary decommissioning:
l
do this, return the device to the condition in which it was transported.
For final decommissioning open the device
l
(p. 70), disassemble the HMI module and turn the back side upwards You are
looking directly at the CPU3 card with the SD card holder(as shown).
Open the SD card holder by sliding the cover down and remove the SD card to
l
prevent unauthorized third parties from accessing the device's data. To pre-
vent data from being recovered even after deletion, we recommend destroying
the SD card.
NOTICE AVIS
Destroying the SD card means, the device will not work any longer.
Détruire la carte SD signifie que l'appareil ne fonctionnera plus.
GasLab Q2
Commissioning and decommissioning
5.1 Storing GasLab Q2
➪  
5.5.1 Opening GasLab Q2
➪ 