Adapting The Factory Parameterization - Honeywell enCore Series Manual

Gas quality analyzer
Hide thumbs Also See for enCore Series:
Table of Contents

Advertisement

6.1.3

Adapting the factory parameterization

In the factory parameterization (default parameter set), usually only calibration gas
properties, automatic calibration intervals and the purging / flushing times for the
gases must be adjusted. For extreme ambient conditions, the setting for the aux-
iliary housing heater or the fan is still required.
The default parameter set automatically switches on the additional heating system
at 0°C ≙ 32°F ≙ 273K with a switching hysteresis of 2°C ≙ 3,6°F ≙ 2K. This sup-
plementary heater function is only available up to a maximum housing temperature
of + 13°C ≙55°F ≙286K.
Other parameters are normally left at their factory-set values and, if necessary, sup-
plemented by customer-specific requirements.
To understand these parameters and to change their values if necessary, please
refer to section
8 Configuration and analysis software enSuite
(p. 171) as well as
➪  
to
enSuite online help
(p. 175)
➪ 
GasLab Q2
User Manual (N )
95 of 315